|
H3: code "other" was selected as the medical device was discarded at facility.Return not possible.A review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: code d12: according to the gore® dryseal flex introducer sheath instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to: vascular trauma (i.E., dissection, rupture, perforation, tear, etc.).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
On (b)(6), 2024, the patient underwent endovascular treatment of a ruptured right internal iliac artery aneurysm using gore® excluder® aaa endoprosthesis and gore® dryseal flex introducer sheath.A 18fr sheath was inserted from the right side.After all planned stent grafts were successfully implanted, angiography for the access route revealed dissection in the right external iliac artery (more distal than the distal end of the stent graft).For repair, a bare metal stent was implanted.The physician reportedly stated as follows: the right external iliac artery was highly tortuous, which made guidewire exchange difficult.It is thought that a very strong force was applied during the guidewire exchange, which may have caused the dissection at that time.
|