• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRUM MEDICAL LTD QUANTUM WORKSTATION (QUANTUM PUMP CONSOLE); HEART LUNG MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRUM MEDICAL LTD QUANTUM WORKSTATION (QUANTUM PUMP CONSOLE); HEART LUNG MACHINE Back to Search Results
Model Number QWS
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2023
Type of Reportable Event Malfunction
Event or Problem Description
Impossible to activate sweep while going on bypass.Sweep flow remained 'blinking' on qvm.No information received if qws patient manager was used as well or not.Perfusionist panicked and weaned back from bypass, has changed the whole quantum console for a backup s5 and finished procedure.No patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM WORKSTATION (QUANTUM PUMP CONSOLE)
Common Device Name
HEART LUNG MACHINE
Manufacturer (Section D)
SPECTRUM MEDICAL LTD
harrier4, meteor business park
cheltenham road east
gloucester, GL2 9 QL
UK  GL2 9QL
Manufacturer (Section G)
SPECTRUM MEDICAL LTD
harrier4, meteor business park
cheltenham road east
gloucester, GL2 9 QL
UK   GL2 9QL
Manufacturer Contact
asia lukuc
harrier4, meteor business park
cheltenham road east
gloucester, GL2 9-QL
UK   GL2 9QL
MDR Report Key21272235
Report Number3006073153-2025-00020
Device Sequence Number3916675
Product Code DTQ
UDI-Device Identifier05060434420008
UDI-Public05060434420008
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K163657
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberQWS
Device Catalogue Number51-000001-00
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/20/2023
Initial Report FDA Received Date01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexUnknown
-
-