• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number AU2110X15B
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2025
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to the manufacturer's factory authorized service center for service.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, a "service required" alarm code indicating an internal battery failure was observed in the device's downloaded event log.The device's power cord, internal battery and main pca board was replaced to address the issue.Additionally, not related to the above issue, the device was alarming for a detachable battery issue.The detachable battery needs to be replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key21273625
Report Number2518422-2025-101503
Device Sequence Number19812386
Product Code CBK
UDI-Device Identifier00606959051980
UDI-Public(01)00606959051980
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberAU2110X15B
Device Catalogue NumberAU2110X15B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2025
Initial Report FDA Received Date01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-