Model Number KD-V411M-0720
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.Under circumstances described above, an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported when the subject device first box was opened, the knife was broken, and when the second box was opened, the knife was also broken.The third one was finally used normally.The issue occurred during sedation for an unknown therapeutic procedure.The procedure was complete with an olympus back up device.There were no reports of patient harm.
 
Event or Problem Description
It was reported when the subject device first box was opened, the knife was broken, and when the second box was opened, the knife was also broken.The third one was finally used normally.The issue occurred during sedation for an unknown therapeutic procedure.The procedure was complete with an olympus back up device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This supplemental emdr is being submitted to correct the manufacturing date: 7/1/2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Common Device Name
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key21274158
Report Number9614641-2025-00098
Device Sequence Number15471756
Product Code KNS
UDI-Device Identifier04953170382628
UDI-Public04953170382628
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K950166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-V411M-0720
Device Lot Number47V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/06/2025
Supplement Date Received by Manufacturer04/08/2025
Initial Report FDA Received Date01/30/2025
Supplement Report FDA Received Date04/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-