| Model Number |
KD-V411M-0720 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.Under circumstances described above, an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported when the subject device first box was opened, the knife was broken, and when the second box was opened, the knife was also broken.The third one was finally used normally.The issue occurred during sedation for an unknown therapeutic procedure.The procedure was complete with an olympus back up device.There were no reports of patient harm.
|
| |
|
Event or Problem Description
|
|
It was reported when the subject device first box was opened, the knife was broken, and when the second box was opened, the knife was also broken.The third one was finally used normally.The issue occurred during sedation for an unknown therapeutic procedure.The procedure was complete with an olympus back up device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental emdr is being submitted to correct the manufacturing date: 7/1/2024.
|
| |
|