| Model Number |
WA22503D |
| Medical Device Problem Codes |
Smoking (1585); Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported, that the hf-resection electrode made large sparks and did not cut.The device was used for bladder resections in two different rooms.It was noted that the device made abnormal noise continuously and smoked and turned black.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the approved final investigation, and update to field d4 and correction to fields h5 and h8.The device history record was unable to be reviewed for this device since the lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.The most likely causes were presumed to be incorrect generator setting, the use of glycine / sorbitol-manitol instead of saline during a bipolar application, or the use of the bipolar electrode with a monopolar working element.These misapplications lead to a malfunction of the electrode.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|