• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RIVERPOINT MEDICAL LLC VELOSORB FAST; BRAIDED FAST ABSORBABLE SURGICAL SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RIVERPOINT MEDICAL LLC VELOSORB FAST; BRAIDED FAST ABSORBABLE SURGICAL SUTURE Back to Search Results
Model Number CV-501G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Wound Dehiscence (1154)
Date of Event 01/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned.Three retained samples from the same lot were used to perform a knot pull tensile test.The retained samples used for the knot pull tensile test met requirements.Production records were reviewed.There were no nonconformities noted with the lot during production.All finished goods testing requirements were met prior to release.There was no evidence that the device failed to meet specifications, and the report could not be substantiated.A cause for the event cannot be established.This report and use of categorical definitions required by fda 3500a does not constitute an admission by riverpoint medical or its employees that riverpoint medical or its employees have caused or contributed to the event described in this report.Riverpoint medical has filed this information to comply with the medical device reporting regulation 21 cfr 803.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
 
Event or Problem Description
According to the reporter, "according to the reporter, six days postoperatively, the patient returned urgently with complete release of the overlock on the two eyes, and it was noted that the suture line did not hold.Attempted repair under local anesthesia proved unsuccessful.The patient had to be taken back under anesthesia generally the following night to redo the two conjunctival sutures with another thread of slower resorption.In addition, the patient extended hospitalization and was admitted to the icu.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELOSORB FAST
Common Device Name
BRAIDED FAST ABSORBABLE SURGICAL SUTURE
Manufacturer (Section D)
RIVERPOINT MEDICAL LLC
825 ne 25th ave
portland OR 97232
Manufacturer (Section G)
RIVERPOINT MEDICAL LLC
825 ne 25th ave
portland OR 97232
Manufacturer Contact
edwin anderson
825 ne 25th ave
portland, OR 97232
MDR Report Key21277104
Report Number3006981798-2025-00002
Device Sequence Number8605894
Product Code GAM
UDI-Device Identifier10884521149458
UDI-Public10884521149458
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K120556
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model NumberCV-501G
Device Lot Number23020810X
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2025
Initial Report FDA Received Date01/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-