Catalog Number 04.033.939S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 01/20/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the broken hardware was removed because the patient stated that he was doing physical therapy at the rehab and heard a pop.The patient was transferred to the er at the hospital and the x-ray images taken showed a broken nail and a fracture.The patient's initial surgery was on (b)(6), 2024.Revision surgery was performed on (b)(6) 2025.The patient was in satisfactory condition following surgery of the removal broken frn nail and implantation of a new femoral nail implant.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information was received and stated that the removal of broken frn nail hardware surgery was successful.The patient was in satisfactory condition following surgery of the removal broken frn nail and implantation of a new femoral nail implant.No other clinical findings found following successfully performed surgery.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: b5, h4.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9 the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1.H3, h6: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device revealed that the 9 / ti cann frn / pf 380 / left - sile was found broken in two parts across the lower locking hole.The fragment were returned for examination.The broken fragment was returned for examination and the fracture surface was examined, this suggest that the fracture occurred post-implantation.Based on the observed evidence, it is reasonable to conclude that the fracture was caused by low-stress, high-cycle fatigue.Additionally it was observed with signs of scratches on the surface which could have been due to implantation and explantation process.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.A dimensional inspection was unable to be performed due to post manufacturing process.A functional evaluation was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the 9 / ti cann frn / pf 380 / left - sile would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: a manufacturing record evaluation was performed for the finished device.Product code: 04.033.939s-us.Lot number: 8949p78.It was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 23 jan 2024.Manufacturing site: jabil bettlach.Expiry date: 31 dec 2033.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9MM / TI CANN FRN / PF 380MM / LEFT - STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach CO
SZ  
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key21277655
Report Number8030965-2025-01130
Device Sequence Number21171088
Product Code HSB
UDI-Device Identifier10886982272799
UDI-Public(01)10886982272799(17)331231(10)8949P78
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K172157
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.033.939S
Device Lot Number8949P78
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/21/2025
Supplement Date Received by Manufacturer01/28/2025
02/13/2025
03/12/2025
Initial Report FDA Received Date01/30/2025
Supplement Report FDA Received Date02/13/2025
02/15/2025
03/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient SexMale
Patient Weight77 KG
-
-