EXACTECH, INC. OPTETRAK ASY,CR CEMENTED FEMORAL, SZ 3,; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
230-02-03 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint effusion (4849)
|
| Date of Event |
11/02/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 9 years post initial left total knee arthroplasty (tka) using cr slope with tibial insert, the patient complained of knee swelling and pain.The surgeon diagnosed the wear of the tibial insert, then revision surgery for the replacement of all knee device was performed.Breakage and wear was observed in the retrieved tibial insert.Also, damage was observed in posterior of tibial tray due to contacting to the femoral component.The surgeon decided to replace all components with a competitor prosthesis.The tibial tray was damaged and found metal powder.The surgeon retrieved and cleaned it appropriately during the surgery.The patient's bone quality was normal and the patient had high activity level.It was also noted that the patient had not worn a knee brace following surgery.The event did not lead to surgical delay/prolongation.The patient was last known to be in stable condition following the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Operative notes and/or medical records were not provided for review of usage/ technique.The revision reported was likely the result of full thickness prosthesis wear of the posterior edge of the tibial insert and subsequent wear between the tibial tray and femoral component secondary to contributions from instability and/or combined tibial slope leading to articulation of the femoral component on the posterior edge of the tibial insert.However, the contributions of instability or component positioning to the sequence of events cannot be confirmed as the devices were not available for evaluation and neither radiographs nor clinical information was provided.
|
| |
|
Search Alerts/Recalls
|
|
|