| Catalog Number |
8065752217 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A physician reported that the phacoemulsification (phaco) tip was broken during the cataract [with intraocular lens (iol) implant] surgery.The surgery was completed by replacing the cassette.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
One opened phacoemulsification (phaco) tip within in a bag was received in a tray with other items.The returned sample was visually inspected and was found to be non-conforming for being broken at the aspiration bypass (ab) hole.The breakage was very jagged, and the wall thickness was observed to be even.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The complaint evaluation confirms the phaco tip was broken.The root cause for the broken phaco tip cannot be determined from this evaluation.The phaco tip visual inspections do not show any manufacturing issues that would cause the broken phaco tip.The exact root cause for the broken phaco tip is unknown; therefore, specific action with regards to this complaint cannot be taken.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance, such as the broken phaco tip exhibited on the returned opened sample, are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|