• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO2, 3CG; PERIPHERALLY INSERTED CENTRAL CATHETER (PICC)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC SOLO2, 3CG; PERIPHERALLY INSERTED CENTRAL CATHETER (PICC) Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
 
Event or Problem Description
It was reported picc line leaking near the insertion site.S/l picc line placed (b)(6) 2024 in our outpatient clinic, the external length was 3cm and a secur-a-cath was used for securement.Patient receiving long term antibiotics (daptomycin) for post-op spine surgery.No patient harm, injury, or complication; no negative outcome.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak is confirmed but the cause is unknown.The product returned for evaluation was a 4fr sl powerpicc solo catheter.The returned product sample was evaluated and a kink impression with a longitudinal split was observed between the 1 cm and 2 cm depth markers on the catheter body.No specific evidence was seen during the investigation which supported a specific root cause for this event.The damage location suggested that catheter securement, access, and/or maintenance techniques may have contributed.This complaint will be recorded for future trending and monitoring purposes.
 
Event or Problem Description
It was reported picc line leaking near the insertion site.S/l picc line placed (b)(6) 2024 in our outpatient clinic, the external length was 3cm and a secur-a-cath was used for securement.Patient receiving long term antibiotics (daptomycin) for post-op spine surgery.No patient harm, injury, or complication; no negative outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO2, 3CG
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER (PICC)
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key21283436
Report Number3006260740-2025-00240
Device Sequence Number7782183
Product Code LJS
UDI-Device Identifier00801741155192
UDI-Public(01)00801741155192(11)241004(17)250930(10)REJW0755
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model NumberN/A
Device Catalogue Number1194108D5
Device Lot NumberREJW0755
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Supplement Date Received by Manufacturer03/06/2025
Initial Report FDA Received Date01/31/2025
Supplement Report FDA Received Date03/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-