| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported picc line leaking near the insertion site.S/l picc line placed (b)(6) 2024 in our outpatient clinic, the external length was 3cm and a secur-a-cath was used for securement.Patient receiving long term antibiotics (daptomycin) for post-op spine surgery.No patient harm, injury, or complication; no negative outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak is confirmed but the cause is unknown.The product returned for evaluation was a 4fr sl powerpicc solo catheter.The returned product sample was evaluated and a kink impression with a longitudinal split was observed between the 1 cm and 2 cm depth markers on the catheter body.No specific evidence was seen during the investigation which supported a specific root cause for this event.The damage location suggested that catheter securement, access, and/or maintenance techniques may have contributed.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|
Event or Problem Description
|
|
It was reported picc line leaking near the insertion site.S/l picc line placed (b)(6) 2024 in our outpatient clinic, the external length was 3cm and a secur-a-cath was used for securement.Patient receiving long term antibiotics (daptomycin) for post-op spine surgery.No patient harm, injury, or complication; no negative outcome.
|
| |
|