• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH CONNECSCR CANN MED; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH CONNECSCR CANN MED; NAIL, FIXATION, BONE Back to Search Results
Catalog Number 03.043.022
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the catalog/model number was not provided, the (01)gtin is not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that there was a problem with drilling a hole for the locking screw due to pressure on the aiming arm system.Surgery was completed successfully with no delay.The surgeon stated they repeatedly came across the nail while drilling.The drilling of the oblique lock was not a problem, the problem was with the connection between the target bar and the nail.It could only be released from the nail with extreme exertion at the end.In the surgeon's opinion, it was initially correctly mounted, which could have contributed to the problem, that they lowered the leg for the distal, oblique locking, that is, the knee stretched and thus certainly brought tension to the connection nail/target bracket.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information was received and states that: a.What approach was used for the tibial nail procedure (suprapatellar, infrapatellar?) infrapatellar.B.Please confirm which hole was being targeted when the problem of hitting the nail occurred (proximal static 1, proximal static 2, dynamic, proximal oblique 1, proximal oblique 2).Proximal static 1, dynamic.C.It was reported that the procedure was successfully completed.Please explain what steps were taken in order to achieve a successful outcome.Proximal oblique holes and free hand locking.D.What was the angle of knee flexion, and did it change during surgery? flexion 45 10, yes.E.Was the compression screw used? no.F.Was the nail over inserted? if yes, how far? no.
 
Event or Problem Description
Additional information was received and stated there were no allegation against the locking screw.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.Investigation summary product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history review (dhr): part number: 03.043.022 lot number: 3101p24 manufacturing site: hägendorf release to warehouse date: 30-nov-2022 a manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECSCR CANN MED
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf CO 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key21283569
Report Number8030965-2025-01167
Device Sequence Number9319983
Product Code JDS
UDI-Device Identifier07612334171030
UDI-Public(01)07612334171030(10)3101P24
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K201336
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.043.022
Device Lot Number3101P24
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/20/2025
Supplement Date Received by Manufacturer02/07/2025
03/24/2025
Initial Report FDA Received Date01/31/2025
Supplement Report FDA Received Date02/13/2025
03/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/29/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - GUIDES/SLEEVES/AIMING.; UNK - SCREWS: LOCKING.
Patient SexUnknown
-
-