ZOLL MEDICAL CORPORATION PROPAQ MD; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number 310-2451312-01
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was evaluated by zoll medical corporation.The customer's report was duplicated and attributed to defective integrated circuit on the processor/bridge/pace board.The board was scrapped after testing.The processor/bridge/pace board will be replaced to resolve the report.The device will be recertified and returned to the customer once the repair estimate has been approved.Analysis for reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that during biomed testing, the device failed self test.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPAQ MD
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key21283941
Report Number1220908-2025-00223
Device Sequence Number20778921
Product Code MKJ
UDI-Device Identifier00847946099633
UDI-Public(01)00847946099633
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P160022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number310-2451312-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2025
Initial Report FDA Received Date01/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1