| Model Number |
26050CA |
| Medical Device Problem Codes |
Leak/Splash (1354); Unintended Electrical Shock (4018)
|
| Health Effect - Clinical Codes |
Electric Shock (2554); Convulsion/Seizure (4406)
|
| Date of Event |
01/07/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation.However, the manufacturer was informed that the device will not be returned for investigation." the event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that leakage occurred for the third time in the use of electric endoscopic instruments in the operation by user, and shocked the patient, but only convulsed during the operation, causing no serious injury, and there was no intervention treatment in the later stage.It was confirmed that the procedure was completed successfully and there was no further impact or consequences.However, due to the reported shock and convulsion of the patient this case is deemed reportable for precautionary reasons.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information:
the product concerned (material-no: 26050ca, lot: zn03) was not returned for investigation.Therefore, it is not possible to investigate the exact cause.It was reported that the product is leaking, but it is unclear where it is leaking.Potential scenarios could be that the shaft was damaged by the user or had previously been repaired by a third-party company.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|