• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS VOLISTA ACCESS; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS VOLISTA ACCESS; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARDVCS209007A
Medical Device Problem Codes Corroded (1131); Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
Getinge became aware of an issue with one of surgical lights - volista access.It was stated that during dismantling of the spring arm corrosion issue was found on various parts of the arms.It was confirmed by photographic evidence the corrosion occured on spring arm and also the paint was peeling from fork.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.
 
Additional Manufacturer Narrative
Event site: (b)(6).Initial report was getinge employee.Getinge became aware of an issue with one of surgical lights - volista access.It was stated that during dismantling of the spring arm corrosion issue was found on various parts of the arms.It was confirmed by photographic evidence the corrosion occured on spring arm and also the paint was peeling from fork.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.Based on the information collected, it was established that when the event occurred, the surgical lights did not meet its specification, since rust and paint peeling could be considered as technical deficiencies, and in this way the device contributed to event.It is unknown if claimed devices were or were not being used for patient treatment or diagnosis when the event took place.According to the subject matter expert at manufacturer¿s, the concentration of chemical products and the stagnation of cleaning agent residues on the disinfected surfaces are the main factors leading to the deterioration of surfaces.To avoid paint degradation and corrosion it is recommended to respect the cleaning instructions, avoiding spraying, high concentrations, prolonged exposure to detergents/disinfectants solutions, and to wipe with a dry cloth and to make sure that no liquid residue is left on the device after cleaning.To reduce the appearance of rust or degradation of the surface, the technical manual or maintenance manuals mention in the preventive maintenance to lubricate some parts of device.To prevent any incident, the user manuals mention to perform daily inspections in order to detect paint defects, impact marks or other damages.Maquet sas recommends to perform corrective maintenance to rectify the default after its detection.Minor paint chips can be repaired with touch up paint, nevertheless the parts impacted by serious damage must be replaced.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLISTA ACCESS
Common Device Name
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key21291026
Report Number9710055-2025-0000046
Device Sequence Number15782935
Product Code FSY
UDI-Device Identifier03700712401351
UDI-Public(01)03700712401351
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K130513
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARDVCS209007A
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2025
Initial Report FDA Received Date02/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2016
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-