| Model Number |
75446540 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 75446545, serial/lot #: (b)(6), udi#: (b)(4); product id: 75446545, serial/lot #: (b)(6), udi#: (b)(4).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having posterior open surgery for degenerative diseases.It was reported that the screws were broken during the procedure.There were no patient symptoms reported.There were no further complications reported regarding the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product analysis: product id# 75446540; lot # h5817359 visual inspection confirmed the screw was returned with the bone screw shank separated from the pedicle head.Macroscopic inspection revealed the multi axial ring has been damaged also due to the bone screw shank pulling loose.The separation appears to be from overload as the root cause.Product analysis: product id# 75446545 lot # h5745459 visual inspection confirmed the screw was returned with the bone screw shank separated from the pedicle head.Macroscopic inspection revealed the multi axial ring has been damaged also due to the bone screw shank pulling loose.The separation appears to be from overload as the root cause.Product analysis: product id# 75446545; lot # h5759850 visual inspection confirmed the screw was returned with the bone screw shank separated from the pedicle head.Macroscopic inspection revealed the multi axial ring has been damaged also due to the bone screw shank pulling loose.The separation appears to be from overload as the root cause.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|