Additional information was provided in a.1., a.2., a.3.A., b.5., b.6., b.3., d.1., d.2., d.3., d.4., d.6.A., d.10.,h.3., h.6.And h.11.The product was not returned.Complaint and product history records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.The lens remains implanted.Each lens is subjected to a 100percentage assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the vision in the right eye has improved to 0.0 logmar unaided for distance and n5 unaided for near following yag and a lasik enhancement procedure in january.The glistenings look symmetrical and now that the refraction has been optimized in the right eye the patient is happy with the vision in that eye.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|