BARD PERIPHERAL VASCULAR, INC. CONQUEST 40; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number CQF7584
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure, when inflating the device, a gap appeared at the proximal end of the balloon where the balloon and shaft meet.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photo and video were provided for review.The investigation of the reported event is currently underway.H11: d2b (lit; dqy).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.One photo was received and reviewed.The photo shows the picture of the sticker from the packaging which matches with the labeling details present.No other anomalies noted.One video was received and reviewed.The video shows the medical professional trying to inflate the balloon.As the video is not very clear, no findings can be determined through the video.No other anomalies noted.Therefore, the investigation is inconclusive for the reported break as no objective evidence was provided for review.A definitive root cause for the reported break could not be determined based upon the provided information.Labeling review: a definitive root cause for the reported break could not be determined based upon the provided information.H10: g3, h6 (method).H11: h6 (result, conclusion).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure, when inflating the device, a gap appeared at the proximal end of the balloon where the balloon and shaft meet.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key21291642
Report Number2020394-2025-00060
Device Sequence Number7775152
Product Code DQY
UDI-Device Identifier00801741060557
UDI-Public(01)00801741060557(17)270430(10)REJS4240
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCQF7584
Device Lot NumberREJS4240
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2025
Supplement Date Received by Manufacturer03/06/2025
Initial Report FDA Received Date02/03/2025
Supplement Report FDA Received Date04/01/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-