Model Number ONB5STF
Medical Device Problem Codes Detachment of Device or Device Component (2907); Insufficient Device Problem Information (3190); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Unintended Radiation Exposure (4565)
Date of Event 01/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 concomitant product: onb5stf, onb5stf versaone opt 5mm std trocar, (lot# unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, in a laparoscopic adnexectomy, the seal of two trocars were damaged.A ct (computed tomography) scan after the surgery revealed that there were metal fragments inside the body.The fragments have a round shape of about 3 mm.
 
Additional Manufacturer Narrative
Additional information: h6 information received shows that this event is a duplicate of another issue reported under regulatory report #.This file will be closed and all further updates processed under the above-referenced report for this event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPORT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21292162
Report Number9612501-2025-00321
Device Sequence Number7585896
Product Code GCJ
UDI-Device Identifier10884521134232
UDI-Public10884521134232
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112349
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberONB5STF
Device Catalogue NumberONB5STF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2025
Supplement Date Received by Manufacturer01/21/2025
Initial Report FDA Received Date02/03/2025
Supplement Report FDA Received Date02/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-