ETHICON ENDO-SURGERY, LLC. CONTOUR CURVED CUTTER STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number CS40G
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 2/3/2025.D4 batch #757c71.B3: only event year known: (b)(6)2024.Investigation summary : the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the cs40g device was received inside of its open package with no apparent damage and with a reload loaded in the device.The reload was received with all staples protruding, the washer uncut and the knife recessed below the reload deck.The drivers were noted to be partially advanced.The device was tested for functionality with a test reload and the device fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.The condition of the reload is consistent with the device being partially activated.As a result of the partial firing the lockout was engaged when the device was reopened.Do not fire the instrument unless the closure trigger is properly latched against the handle.The firing trigger must be pulled back completely against the closure trigger to properly fire the instrument.In addition, if the firing sequence is not complete, you could deploy the staples without cutting the washer and forming the staples.Please reference instructions for use for additional information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 757c71, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device cs40g lot number a9ek0l and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: 1.Could you please clarify if there were any patient consequences? 2.Could you please clarify if there was any change in the post-operative care of the patient as a result of the event?.
 
Event or Problem Description
It was reported that during an unknown procedure the wire of the device doesn't work, so they couldn't fire it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR CURVED CUTTER STAPLER
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key21292440
Report Number3005075853-2025-00876
Device Sequence Number15764179
Product Code GDW
UDI-Device Identifier10705036012405
UDI-Public(01)10705036012405(17)281130(10)A9EK0L
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K040038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCS40G
Device Lot NumberA9EK0L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2025
Initial Report FDA Received Date02/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-