• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN MINI ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN MINI ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number D023
Medical Device Problem Codes Pacing Problem (1439); Inaccurate Synchronization (1609); Under-Sensing (1661)
Health Effect - Clinical Code Arrhythmia (1721)
Date of Event 01/09/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) stored two episodes that were analyzed.The first episode showed a non-sustained ventricular tachycardia where an atrial pace blanked the ventricular intrinsic deflection.This occurred due to a device algorithm that caused premature atrial pacing, then the device paced in the ventricle, and this appeared to have induced a ventricular arrhythmia.At the end of the episode the rate fell below the therapy zone.Some minutes later a new ventricular episode started where appropriate anti-tachycardia pacing, and shocks were delivered.Reprogramming suggestions were discussed with the caller.The icd remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN MINI ICD DR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21292805
Report Number2124215-2025-05318
Device Sequence Number12699371
Product Code LWS
UDI-Device Identifier00802526533921
UDI-Public00802526533921
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/06/2024
Device Model NumberD023
Device Catalogue NumberD023
Device Lot Number255134
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2025
Initial Report FDA Received Date02/03/2025
Date Device Manufactured01/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age86 YR
Patient SexMale
-
-