• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRUM AB OPERA IMPLANT SYSTEM; AXOR II

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRUM AB OPERA IMPLANT SYSTEM; AXOR II Back to Search Results
Model Number 1288
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
(b)(6) 2025: integrum has been informed that axor ii unit (b)(6) has fallen off from the abutment.No health consequences or patient injury reported.Manufacturing investigation performed; batch documentation reviewed.No deviations found; the device has been manufactured according to specification.(b)(6) 2025: technical investigation of unit (b)(6) performed.During the investigation the functions were tested and components inspected.The reported issue could not be replicated.The clamps showed marks which indicates that the abutment may not have been fully inserted at some point.In addition, several components were damaged and incredibly dirty due to poor maintenance.During service, worn components were replaced.The device was thoroughly cleaned and functionally tested according to specification with approved results.
 
Event or Problem Description
(b)(6) 2025: integrum has been informed that axor ii unit(b)(4).Has fallen off from the abutment.No health consequences or patient injury reported.Manufacturing investigation performed, batch documentation reviewed.No deviations found, the device has been manufactured according to specification.(b)(6) 2025: technical investigation of unit (b)(4).Performed.During the investigation the functions were tested and components inspected.The reported issue could not be replicaded.The clamps showed marks which indicates that the abutment may not have been fully inserted at some point.In addition, several components were damaged and incredibly dirty due to poor maintenance.During service, worn components were replaced.The device was thoroughly cleaned and functionally tested according to specification with approved results.(b)(6) 2025: a feedback report sent to the customer highlight the importance of donning the axor ii according to the ifu to ensure a stable connection and prevent damage or wear to the clamps.The importance to move the leg/tap the foot on the floor, tighten further and check the attachment.As stated in the ifu, the device shall not be used if a stable connection cannot be achieved.The importance of cleaning the axor ii in accordance with the ifu and patient information.Failure to adhere to the instructions may affect the axor ii's function.Correct cleaning is also important prior to shipment for service.Failure to adhere to the instructions can compromise the safety of integrum staff.A reminder to send the device to integrum for annual service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPERA IMPLANT SYSTEM
Common Device Name
AXOR II
Manufacturer (Section D)
INTEGRUM AB
gemenskapens gata 9
molndal, 43153
SW  43153
Manufacturer (Section G)
INTEGRUM
gemenskapens gata 9
molndal, 43153
SW   43153
Manufacturer Contact
susanne börjesson
gemenskapens gata 9
molndal, 43153
SW   43153
MDR Report Key21294177
Report Number3011386779-2025-00216
Device Sequence Number21143448
Product Code PJY
UDI-Device Identifier07340152100634
UDI-Public07340152100634
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P190009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial,Followup
Report Date (Section B) 02/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1288
Device Lot Number56381
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2025
Supplement Date Received by Manufacturer01/21/2025
Initial Report FDA Received Date02/03/2025
Supplement Report FDA Received Date02/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-