| Model Number |
305U227 |
| Medical Device Problem Codes |
Gradient Increase (1270); Mechanics Altered (2984); Insufficient Device Problem Information (3190); Thickening of Material (4056); Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Aortic Valve Stenosis (1717); Insufficient Information (4580)
|
| Date of Event |
01/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Medtronic received information that 10 years and 3 months post implant of this aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was not reported.No intervention or adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
B1, b2, b5, h1 and imf (annex f) codes were updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that on (b)(6) 2025, the patient successfully had a 29mm transcatheter valve-in-valve replacement.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated data: b.5: event description h.6: coding medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received that reported the reason for replacment as a significant reduction in leaflet mobility, a low fl ow/low gradient of of 14mmhg, aortic stenosis, a peak velocity of 2.5m/s and a reduced ejection fraction (ef) of 40-45%.No additional adverse patient effects were reported.
|
| |
|