|
Device 8 of 11.Based on the available information, this event is deemed to be a reportable malfunction.Photograph, video and/or physical sample evaluation: there are photographs associated with this case and in these, the reported defect can be seen.No unused return sample was expected.Batch record revision results: lot: 3g03485, order: (b)(4), material: 1000642, was manufactured on 01/jul/2023 in the convex 1 piece (pc) manufacturing line, with a total of (b)(4) market units (mkus).The complaint investigator performed a batch record review on 30/jan/2025 and it was confirmed that the applicable procedures were followed, the components for assembly were correct as per bill of materials (bom) and the equipment settings were the applicable ones.The batch record review supports that there were no discrepancies related to the issue reported.Historical complaints review: on 30/jan/2025, complaints investigator ran a query in database in order to verify the complaints reported for the 3g03485 lot for the malfunction ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿ defect and no additional complaint was identified.Historical nonconformance review: on 30/jan/2025, complaints investigator ran a query in database looking for any in process nonconformance / corrective action / preventive actions (capa)(s) associated to the malfunction ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿ defect for the lot number: 3g03485 and as result, no nonconformance / capa (s) for this malfunction were generated during the manufacturing process of the referenced lots.Defect rate analysis: there have only been 01 defective part confirmed to date from a lot size of (b)(4) products.This represents a defect rate of only (b)(4), which is well within an appropriate acceptable quality level (aql) for this defect which should be 0.25% based on our standard operating procedure (sop).In addition, all of the in-process testing on this lot did not find a single defective unit, which confirms that the lot is unlikely to breach an aql of 0.25.To date, it is well within our accepted aql level for this type of failure mode or defect.Conclusions: as photographs were available for this complaint issue, they were evaluated and as a result, the reported malfunction for the observed product was confirmed, however, after performing the defect rate analysis for this issue, it was observed that the quantity of reportedly affected products is within the appropriate aql for this defect.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092, manufacturing site: 9618003.
|