• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIAC SCIENCE CORPORATION POWERHEART G5 KIT, AUTO, US ENG/ESP LA; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIAC SCIENCE CORPORATION POWERHEART G5 KIT, AUTO, US ENG/ESP LA; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number G5A-80A
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including daily/weekly/monthly self-tests, on/off current, patient and circuit impedance testing, and shock testing without duplicating the report.An internal inspection found no discrepancies.The g5 main board will be replaced as a precaution.The board was scrapped after testing.The device will be recertified and returned to the customer once the repair estimate has been approved.Analysis of reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that during functional testing, the device displayed an "ecg fault -501" error message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERHEART G5 KIT, AUTO, US ENG/ESP LA
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
CARDIAC SCIENCE CORPORATION
500 burdick parkway
deerfield WI 53531 9692
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key21295612
Report Number2112020-2025-00049
Device Sequence Number7775265
Product Code MKJ
UDI-Device Identifier00812394020546
UDI-Public(01)00812394020546
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P160033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberG5A-80A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2025
Initial Report FDA Received Date02/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-