BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE; IMPLANTABLE LEAD Back to Search Results
Model Number 0147
Medical Device Problem Codes High impedance (1291); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this right ventricular (rv) lead had a rise in the pacing and shock impedances then a sharp drop.Prior to the drop the impedance measurements were over 2000 ohms.Technical services (ts) noted no oversensing.The representative is planning to check if the patient had a procedure at the time of the event.This rv lead remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key21296214
Report Number2124215-2025-05468
Device Sequence Number9208140
Product Code LWS
UDI-Device Identifier00802526085789
UDI-Public00802526085789
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/25/2009
Device Model Number0147
Device Catalogue Number0147
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2025
Initial Report FDA Received Date02/03/2025
Date Device Manufactured11/25/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexMale
-
-