| Model Number |
383074 |
| Medical Device Problem Codes |
Under-Sensing (1661); Decreased Sensitivity (2534)
|
| Health Effect - Clinical Codes |
Tachycardia (2095); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2025
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited undersensing prior to detection of a ventricular fibrillation (vf) ep isode.The rv lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Continued from d10: 694045 lead, 6945-65 lead implanted: (b)(6) 1999.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that reprogramming was done.A little over a month later, the patient passed away.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that before the patient passed, the rv lead exhibited a drop in the r wave measurement.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected: b5 and h6 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited undersensing prior to detection of a ventricular fibrillation (vf) ep isode.The rv lead remains in use. no patient complications have been reported as a result of this event.It was further reported that reprogramming was done.A little over a month later, the patient passed away.It was further reported that before the patient passed, the rv lead exhibited a drop in the r wave measurement.It was also reported that undersensing of ventricular tachycardia (vt) was observed in the patient's last remote transmission.
|
| |
|