| Model Number |
VXSK0022 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Device-Device Incompatibility (2919); Material Deformation (2976); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was mentioned that versacross access solution was selected for watchman procedure.It was mentioned that when mechanical guidewire was inserted with dilator over the wire, the guidewire got stuck into the dilator and the dilator cannot be removed.Thus, the dilator and mechanical guidewire were removed, and the procedure was completed successfully using versacross rf wire.No patient complications were reported.The device is not expected to be return for analysis.The guidewire got kinked and was stuck into the dilator, which made it difficult to remove the guidewire from the dilator.When the wire got stuck, the distal tip of the wire was exposed past to the distal end of the dilator.
|
| |
|