• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. CLARIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. CLARIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTMA1Q1
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/22/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: 694758 lead implanted: (b)(6) 2009, 429888 lead implanted: (b)(6) 2014.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited premature battery depletion.The crt-d was explanted and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.The device was returned and analyzed.The device met 97.7% of expected longevity.Without the history of the programmed settings throughout its service life, there is no way to determine why the longevity did not match the predicted model.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI QUAD CRT-D SURESCAN
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key21300045
Report Number3004209178-2025-01963
Device Sequence Number8169460
Product Code NIK
UDI-Device Identifier00643169837669
UDI-Public00643169837669
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/28/2021
Device Model NumberDTMA1Q1
Device Catalogue NumberDTMA1Q1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2025
Supplement Date Received by Manufacturer03/27/2025
Initial Report FDA Received Date02/03/2025
Supplement Report FDA Received Date04/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-45 LEAD.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age54 YR
Patient SexMale
-
-