• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY LESSER TMT FIXATION PACK; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY LESSER TMT FIXATION PACK; PLATE Back to Search Results
Model Number SK28
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after an initial bunion surgery on (b)(6) 2024, a broken plate at the third tmt was discovered via a radiographic image from a 6-month post-op follow-up.No other patient outcomes were reported as a result of this event.Although there has been no injury reported and no information has been received indicating this malfunction has resulted in a revision surgery to date, there have been similar events where this malfunction has resulted in a revision surgery.Therefore, an mdr will be filed to ensure full compliance with 21 cfr part 803.
 
Additional Manufacturer Narrative
It was reported that after an initial bunion surgery on (b)(6) 2024, a broken plate at the third tmt was discovered via a radiographic image from a 6-month post-op follow-up.No devices were returned for evaluation as the hardware remains implanted.The device history records were reviewed and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.A number of factors could have contributed to what was reported and although the most likely cause cannot be determined based on the available information, the broken plate was likely subjected to excessive bending stresses which resulted in its breakage.The company will supplement the mdr as necessary and appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPIPLASTY LESSER TMT FIXATION PACK
Common Device Name
PLATE
Manufacturer (Section D)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer (Section G)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer Contact
renae ginter
100 palmetto park place
ponte vedra, FL 
9043735940
MDR Report Key21300706
Report Number3011623994-2025-00011
Device Sequence Number15683936
Product Code HRS
UDI-Device Identifier00853114006204
UDI-Public(01)00853114006204(17)280510(10)300427053
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K220136
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSK28
Device Lot Number300427053
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Initial Report FDA Received Date02/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexMale
-
-