| Model Number |
2090 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the programmer shut down.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: analysis was able to confirm the customer comment that the programmer shut down.The programmer powered on initially but shut down and could not power on anymore.The power supply was replaced.It was noted that the power cord bay assembly was broken.During software fast load, the programmer did not respond to the stylus.The programmer was decommissioned as it was no longer serviceable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|