RESPIRONICS, INC. DREAMSTATION BIPAP ST30 H; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
|
Back to Search Results |
|
| Model Number |
DSX1030H11 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The patient alleges visualization of particles in device.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation has been completed.H11 should be: the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The patient alleges visualization of particles in device.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The device was returned to a third-party service center.During the device evaluation, the technician confirmed no foam particles in the air path and no secondary findings was observed.The device was scrapped.During evaluation there was no error codes observed.The third-party service center concludes that they couldn't confirm the customer's allegation.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.Due date has been updated.In box g: date received by mfg has been updated.In box h6: evaluation method code grid, evaluation results code grid and conclusion code grid have been updated.
|
| |
|
Search Alerts/Recalls
|
|
|