• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFCSIG +/-2FEMOFFSET STMBLT; SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFCSIG +/-2FEMOFFSET STMBLT; SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY Back to Search Results
Catalog Number 960771
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pocket Erosion (2013); Unspecified Tissue Injury (4559); Insufficient Information (4580); Bone Notching (4890)
Date of Event 11/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d4: udi: as the catalog/model number was not provided, the (01) gtin is not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that in a partially coupled, cemented revision prosthesis, a fracture of the femoral stem and the connecting screw occurred without trauma in the connection area to the femoral shield.
 
Event or Problem Description
Additional information received describes a revision procedure requiring extensive debridement.Extensive bone erosion was noted on the medial and lateral femoral condyles as well as along the medial tibial plateau requiring bone grafting to correct.Granulation tissue indicative of loosening was noted.The femoral and tibial sides were noted to be loose - interface not identified.The patella was removed and underneath small cysts were identified and removed before replacing.This is the right side.The date of operation for the first implantation was on (b)(6) 2008, without complications.The revision went smoothly.The connecting screw between the femoral shield and the femoral shaft is affected.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: b5, d6a, d10, g4, h4 and h6 (health effect - clinical code).Corrected: d1, d2, d4 (lot, catalog, expiration date, udi).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: user report received bfarm case 1652/25 in the above-mentioned partially coupled, cemented revision prosthesis, a fracture of the femoral stem and the connecting screw occurred without trauma.Femoral stem and the connecting screw in the connection area to the femoral shield the product was not returned to j&j medtech orthopaedics; however, photos were provided for review.The photo investigation revealed that the distal threaded portion of the pfcsig +/-2femoffset stmblt has fractured.A manufacturing record evaluation was performed for the finished device [960771/ 4570771d], and no non-conformances or manufacturing irregularities were identified.Although a definitive cause for device fracture is not established, one contributing factor appears to be loosening at the interface between the cement and bone.The specific manufacturer of the cement remains unknown; however, it is evident that an excess amount of cement was used to fill bone voids in the anterior portion of the distal femoral head.This likely resulted in inadequate fixation between the cement and the bone.Insufficient fixation can lead to micromotion, which may increase stress at the junction of the stem bolt and the stem.This can result in fatigue cracks and ultimately catastrophic failure.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the pfcsig +/-2femoffset stmblt would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : 1) quantity manufactured: (b)(4).2) date of manufacture: 7-dec-2008.3) any anomalies or deviations identified in dhr: none.4) expiry date: 30-nov-2012.5) ifu reference: ifu-0902-00-252.Device history review : a manufacturing record evaluation was performed for the finished device [960771 / 4570771d], and no non-conformances or manufacturing irregularities were identified.Corrected: d4 (expiry date, primary udi number), h4, h6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFCSIG +/-2FEMOFFSET STMBLT
Common Device Name
SIGMA REVISION IMPLANT : KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21302617
Report Number1818910-2025-02010
Device Sequence Number12987559
Product Code JWH
UDI-Device Identifier10603295233909
UDI-Public(01)10603295233909(17)121130(10)4570771D
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K952830
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2012
Device Catalogue Number960771
Device Lot Number4570771D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/20/2025
Supplement Date Received by Manufacturer02/14/2025
06/18/2025
Initial Report FDA Received Date02/04/2025
Supplement Report FDA Received Date02/27/2025
06/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/07/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BIOSTOP G CEM RESTR 12MM; PFC SIGMA FEM POST AUG SZ3 4MM; PFC SIGMA TIB TRAY CEM SZ3; PFC*SIGMA DIS AUG 4MM,SZ3,RGHT; PFC*SIGMA DIS AUG 4MM,SZ3,RGHT; PFC*SIGMA TC3 FEM RT SZ3; PFC*SIGMA TC3 INS 22.5MM, SZ 3; PFC*SIGMA/OV/DOME PAT 3PEG,38; PFCSIG +/-2FEMOFFSET STMBLT; PFCSIG CEM FEM STEM 5DG13X90MM; SIG TIB CEM STM 13X60 2/2.5/3
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-