• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® PEDIATRIC FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® PEDIATRIC FOLEY CATHETER Back to Search Results
Catalog Number 0165PL10
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.The product was used for urological care.It was unknown whether the product had caused the reported failure.However, the potential root cause for this failure mode could be the catheters was contact with sharp object or exposure to petrolatum based products.No actions can be taken at this time since a root cause was not identified.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: structure & composition: the product include three series: bardex, bardic and bardia.The composition of each series is as follows: bardex, bardic and bardia bi-lumen series are composed of catheter body, inflation cone interface, balloon and valve.Bardia tri-lumen is composed to tube body, inflation cone interface, deflation cone interface, flush cone interface, balloon and valve.The product is with silicone coating.The material of catheter body, inflation conical interface, deflation conical interface, flushing conical interface, balloon is natural latex.The material of valve is polypropylene.Sterilized by gamma radiation.Disposable.Scope of application: foley catheter is intended for use in the drainage of urine from the bladder of children and adults.Precaution: this product contains natural rubber latex which may cause allergic reaction.Sterile unless package has been opened or damage.Warning: do not use petroleum substrate lubricants with the latex- based urethral catheters such as petroleum jelly and label liquid paraffin, which will damage latex and may burst balloon.Water-soluble lubricant can be used.Valve type: use luer slip syringe.Do not use needle.To deflate catheter balloon: gentle insert a syringe in the catheter valve.Never use excessive force to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen.If permitted by hospital practical, the valve may be cut off.If this fails, please contact an adequately trained professional for assistance, as directed by hospital practical.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Storage: catheters need to be stored at room temperature and keep away from the direct light and preferably stored in the original packing box.Do not use the product of have latex allergy.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the when the mid shift nurse was patrolling the ward, they discovered that the foley catheter in bed 1409 had fallen out.They checked immediately and found that the balloon of the catheter was damaged and the catheter had fallen out on its own, so they reinserted the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® PEDIATRIC FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key21303428
Report Number1018233-2025-00548
Device Sequence Number8922277
Product Code EZC
UDI-Device Identifier00801741017674
UDI-Public(01)00801741017674(11)230824(17)280428(10)MYHT0858
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0165PL10
Device Lot NumberMYHT0858
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2025
Initial Report FDA Received Date02/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-