Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 5484306, serial/lot #: (b)(6), udi#: (b)(4) ; product id: 5484306, serial/lot #: (b)(6), udi#: (b)(4); product id: 5484306, serial/lot #:(b)(6), udi#: (b)(4); product id: 5584111, serial/lot #: (b)(6), udi#: (b)(4).H3: product analysis: part # 5484306; lot # (b)(6) - visual inspection confirmed the torx tip of instrument has been damaged consistent with interface during usage.Optical examination revealed the torx tip edges have been rolled over and deformed.This type of damage is consistent with torsional overload.Part # 5484306; lot # (b)(6) - visual inspection confirmed the torx tip of instrument has been damaged consistent with interface during usage.Optical examination revealed the torx tip edges have been rolled over and deformed.This type of damage is consistent with torsional overload.Part # 5484306; lot # (b)(6) - visual inspection confirmed the torx tip of instrument has been damaged consistent with interface during usage.Optical examination revealed the torx tip edges have been rolled over and deformed.This type of damage is consistent with torsional overload.Part # 5484306; lot # (b)(6) - visual inspection confirmed the torx tip of instrument has been damaged consistent with interface during usage.Optical examination revealed the torx tip edges have been rolled over and deformed.This type of damage is consistent with torsional overload.Part # 5584111; lot # (b)(6)- visual inspection confirmed the torx tip of instrument has been damaged and the attachment pin to the internal bushing has broken consistent with interface during usage.Optical examination revealed the torx tip edges have been rolled over and deformed.This type of damage is consistent with torsional overload.This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|