Catalog Number 180-01-58
Medical Device Problem Codes Mechanical Problem (1384); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Scar Tissue (2060); Osteolysis (2377); Implant Pain (4561); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/11/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
1038671-2023-01224 d10: concomitant devices 122-65-25 - 6.5mm acetabular bone screw 25mm (b)(6).122-65-25 - 6.5mm acetabular bone screw 25mm (b)(6) 170-36-00 - biolox delta femoral head 36mm od, +0mm (b)(6).188-01-09 - wedge plasma x/o sz 9 (b)(6), multiple mdr reports were filed for this event, please see associated report: 1038671-2023-01224.The most likely underlying cause for the revision reported is prosthesis wear, loss of range of motion, and acetabular loosening.And a combination of risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed from the reported information and the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported via legal documentation that a patient had an initial right hip arthroplasty and then approximately 7 years, 11 months later the patient experienced a revision surgery.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV CROWN CUP CLSTR HOLE 58MM GROUP 3
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21303906
Report Number1038671-2025-00634
Device Sequence Number21051848
Product Code JDI
UDI-Device Identifier10885862037046
UDI-Public10885862037046
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/04/2019
Device Catalogue Number180-01-58
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2023
Initial Report FDA Received Date02/04/2025
Date Device Manufactured11/07/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexMale
-
-