| Model Number |
H938751 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
I was reported that an exactamix vented micro-volume inlet had particulate matter (pm).The pm was further described as ¿large brown particles¿.This was noticed during priming.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: particulate matter was identified in the final bag, not in the inlet specifically.However, the customer indicated the issue was due to non-baxter tubing; therefore, the baxter product is no longer suspect in this event.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|