• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number PXT231412
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The following information was reported to gore: on (b)(6) 2008, this patient underwent endovascular treatment for an abdominal aortic aneurysm using gore® excluder® aaa endoprostheses.1x trunk ipsilateral leg and total 4x contralateral legs were implanted, extending to both common iliac arteries.On an unknown date, expansion of both common iliac arteries was observed.On (b)(6) 2025, a reintervention was performed.Both internal iliac arteries were coil-embolized, and additional stent grafts were placed in each external iliac arteries.Furthermore, a bird-beak was observed at the proximal end of the trunk ipsilateral leg.Although no endoleak was observed, an additional stent graft was placed proximally as a preventive measure.The physician¿s comment: ¿as we do not have surgical records, it is unclear whether the initial placement of stent grafts was suboptimal or if there was an error in size selection.Although more than 16 years had passed since the initial placement, no damage or deterioration of the device was observed.¿.
 
Additional Manufacturer Narrative
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: code d12: according to the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and / or require intervention or additional intraoperative procedure time include, but are not limited to: endoleak.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
eimi ueda
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key21304023
Report Number3007284313-2025-03736
Device Sequence Number8169597
Product Code MIH
UDI-Device Identifier00733132610136
UDI-Public(01)00733132610136(17)110731
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2011
Device Catalogue NumberPXT231412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2025
Initial Report FDA Received Date02/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/05/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
Patient SexMale
-
-