| Model Number |
ARTHREX CHONDRAL DART, 18MM (5-BOX) |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a chondral flap repair surgery the chondral dart broke.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.Upon visual inspection, it was noted that the chondral dart 18mm was broken off.The piece returned measured 5.89 of.709 ±.010 per drawing (b)(4).Functional testing was not performed due to the damage to the device.The method of bone preparation and the quality of the bone encountered were not specified.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, misalignment insertion, and/or prying/leveraging of the device during insertion.
|
| |
|
Search Alerts/Recalls
|