BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24690
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the severely tortuous and severely calcified superficial femoral artery.A 6.0mmx220mmx150cm (4f) sterling balloon catheter was advanced for dilation.During fifth inflation at 8 atmospheres for 20 seconds, the balloon ruptured at the middle part.The device was removed using normal method without issue and the procedure was completed with a different device.No complications were reported, and the patient was in good condition post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
6128757234
MDR Report Key21305503
Report Number2124215-2025-05700
Device Sequence Number21044959
Product Code LIT
UDI-Device Identifier08714729859659
UDI-Public08714729859659
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141150
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24690
Device Catalogue Number24690
Device Lot Number0034837301
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2025
Initial Report FDA Received Date02/04/2025
Date Device Manufactured09/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-