MAUDE Adverse Event Report: ETHICON INC. ETHICON SECURESTRAP; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. ETHICON SECURESTRAP; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number STRAP12
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested and the following was provided: - were there any patient consequences? the patient's condition is stable.- was there any associated medical or surgical intervention to treat any of the patient consequences? if so, please clarify.The bleeding site was stopped by an electric knife.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
 
Event or Problem Description
It was reported that a patient underwent a laparoscopic unknown surgery on an unknown date and absorbable straps were used.The device (100tp8) did not deploy straps at the 1st firing, and half of the clip was ejected at first the 2nd firing.One clip popped out when an air shot with the device outside of the surgical field.The device (100esa) also did not deploy straps at the 1st firing, and the clip was adhering to the upper abdominal wall.It could not be fixed to the abdominal wall at the 1st firing, but it was confirmed that there was bleeding from the site where it was hit.The bleeding site was stopped by an electric knife.The grasping force of both devices was higher than expected.A competitive device was used as a replacement.Two devices will be returning.No further information is available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h4.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h3, h6.H3 investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the strap12 device was returned with no apparent damage in the external components and a petri dish with one(1) straps inside.In addition, a foil pouch was received.Upon cycling, the instrument was noted to be empty and locked out.For fire testing the trigger was unblocked and functional test was conducted successfully, the device worked as intended.The device was disassembled and no straps were found inside cannula shaft.No abnormalities were noted on internal device components.The provided device was fully dispensed.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.The reported complaint could not be confirmed.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHICON SECURESTRAP
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key21307120
Report Number2210968-2025-01348
Device Sequence Number20838134
Product Code GDW
UDI-Device Identifier10705031132986
UDI-Public(01)10705031132986(17)260331(10)100ESA
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093845
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Catalogue NumberSTRAP12
Device Lot Number100ESA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Supplement Date Received by Manufacturer02/05/2025
02/05/2025
Initial Report FDA Received Date02/04/2025
Supplement Report FDA Received Date02/28/2025
02/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-