|
MENTOR TEXAS MENTOR SMOOTH ROUND SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3501480 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
09/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Female patient who underwent breast reconstruction revision using a mentor smooth round spectrum implant (475cc on the left) and an unknown allergan sclp saline implant on the right experienced a left-sided deflation, and right sided asymmetry issue post-implantation.In response to this complication, the patient underwent a bilateral periprosthetic capsulectomy and bilateral replacements.The left breast implant was replaced with catalog number smpb450, serial number (b)(6), while the right breast implant was replaced with catalog number smpb480, serial number (b)(6).Additionally, the procedure included bilateral fat grafting to the reconstructed breasts and bilateral mastopexy, which was performed on (b)(6) 2024.This report specifically pertains to the left side of the patient's reconstruction.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on february 20, 2025: the product was returned to mentor for evaluation.Mentor conducted visual inspection and microscopic examination of the returned device.Visual analysis of the returned device revealed that the spec smooth rnd 475cc breast implant had a crease/fold on the posterior view.In addition, a tear was found within the crease/fold, measuring approximately 0.1 cm.A microscopic examination was performed, and no instrument damage was observed at the edges of the rupture.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a deflation in a later time.Breast implants may also simply wear out over time.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|