MEDTRONIC IRELAND IN.PACT AV ACCESS; DRUG-COATED PERIPHERAL TRANSLUMINAL ANGIOPLAST Back to Search Results
Catalog Number IAV08004008P
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 12/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
During the index procedure two in.Pact av access balloons were used to treat the right arm venous outflow.One in.Pact av access was used later to treat right arm.Approximately 23 months post procedure patient suffered thrombosis in arteriovenous fistula.Physician unable to complete previous hemodialysis due to clot.Angiography performed demonstrated in-stent thrombosis.Thrombectomy was performed using a cleaner device.Fogarty balloon used to pull arterial plug.Angioplasty using 6 x 40 mustang was performed near brachial anastomosis.Repeat angioplasty demonstrated opacification throughout inflow without residual stenosis.Successful de-clot of right upper extremity av brachio-axillary graft.Target lesion, venous outflow, involved in procedure.
 
Additional Manufacturer Narrative
Update 27 feb, 03, 17 mar 2025: during the index procedure the right arm venous outflow in-graft segment was treated.The right arm venous in-graft segment was also treated in the subsequent procedure.Ae term updated to: thrombosis in arteriovenous graft.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IN.PACT AV ACCESS
Common Device Name
DRUG-COATED PERIPHERAL TRANSLUMINAL ANGIOPLAST
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key21307734
Report Number9612164-2025-00584
Device Sequence Number12804044
Product Code ONU
UDI-Device Identifier00763000232641
UDI-Public00763000232641
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P190008
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/03/2024
Device Catalogue NumberIAV08004008P
Device Lot Number0010589846
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Supplement Date Received by Manufacturer02/27/2025
Initial Report FDA Received Date02/04/2025
Supplement Report FDA Received Date03/25/2025
Date Device Manufactured03/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight64 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-