| Catalog Number |
0380544100 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device breaking during the procedure.All broken pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device breaking during the procedure.All broken pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Three attempts has been made to retrieve the device for evaluation, and it is yet to be received in-house.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: shaver tip broke off probable root cause: inadequate window profile inadequate welding process improper material strength inadequate size selected for the application material brittleness excessive force applied by the user incorrect rfid tag the reported failure mode will be monitored for future reoccurrence.Added initial reporter phone number.
|
| |
|