EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 6 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-35-6011
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Scar Tissue (2060); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Joint effusion (4849)
Date of Event 11/08/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately 93 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, ambulation difficulties, discomfort, pain, joint effusion, scar tissue, and osteolysis.Initial surgery and revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
D10: (b)(6) - 02-010-01-0260 - logic femoral ps cem left sz 6; (b)(6) - 02-010-01-0360 - logic femoral ps cem right sz 6; (b)(6) - 02-012-35-6009 - logic tibia ps mod insrt sz 6 9mm; (b)(6) - 02-012-45-6050 - lgc tibial fit tray cem sz 6f / 5t; (b)(6) - 200-02-41 - three peg patella 41mm; (b)(6) - 200-02-41 - three peg patella 41mm; (b)(6) - 02-012-45-6050 - lgc tibial fit tray cem sz 6f / 5t.The product associated with the reported event is within the scope of recall [z-0021-2022]; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2022-01530.The reason for the revision reported cannot be confirmed from the information provided but may be the result of femoral loosening and prosthesis wear or due to inclusion of the polyethylene in the packaging recall.Failure of the cement used to secure the femoral component to the femoral bone, may have lead to loosening at the cement-implant interface; however, this cannot be confirmed.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 6 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21312528
Report Number1038671-2025-00654
Device Sequence Number15472818
Product Code JWH
UDI-Device Identifier10885862001979
UDI-Public10885862001979
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K093360
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/06/2020
Device Catalogue Number02-012-35-6011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2023
Initial Report FDA Received Date02/05/2025
Date Device Manufactured01/09/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age74 YR
Patient SexMale
-
-