Catalog Number 71675383
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/18/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, after an internal fixation surgery, had been performed on (b)(6) 2025, the intertan 10s 10mm x 18cm 125d broke at the level of the distal screw hole.This adverse event was treated by a revision surgery on (b)(6) 2025, in which a intertan 10s 10mm x 18cm 125d was exchanged.Current health status of patient is unknown.No further complications were reported.
 
Additional Manufacturer Narrative
Additional information: h6 (type of investigation), h10 (roi).H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical investigation found that using a short intertan nail for a complex fracture likely caused the issue, as it's not recommended for such cases.According to the instructions for use, it is noted that the ¿short intertrochanteric nail is contraindicated for complex intertrochanteric and femoral neck fractures.¿ the patient's history of falls and early weight-bearing might have also contributed, although the patient claims the fracture happened while in bed.The x-ray shows a fracture near the nail, but without comparison images, it's unclear if a periprosthetic fracture occurred or if the bone healed properly.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed part number over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for intramedullary nail system revealed that in possible adverse effects that cracking or fracture of the implant may occur.Preoperative planning section states that proper type and size of implant must be selected to ensure effective treatment of patients considering factors such as patient¿s size, strength, skeletal characteristics, skeletal health, and general health.Overweight or musculoskeletal deficient or unhealthy patients may create greater loads on implants that may lead to breakage or other failure of the implants.Additionally, postoperative care section warns that early weight bearing substantially increases implant loading and increases the risk of breaking the device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According with inspection drawing, the final inspection includes the verification of part configuration per print.Also, per material specification the quality and manufacture of standard grade titanium-6 aluminum-4 vanadium alloy shall be controlled.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.H6 (investigation conclusions).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 10S 10MM X 18CM 125D
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key21313114
Report Number1020279-2025-00223
Device Sequence Number15472831
Product Code JDS
UDI-Device Identifier00885556105696
UDI-Public885556105696
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K040212
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71675383
Device Lot Number23CM07771
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2025
Supplement Date Received by Manufacturer03/07/2025
Initial Report FDA Received Date02/05/2025
Supplement Report FDA Received Date03/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/14/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age62 YR
Patient SexMale
Patient Weight88 KG
-
-