• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ PMIC/ID-106; SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO. (SPARKS) BD PHOENIX¿ PMIC/ID-106; SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION Back to Search Results
Catalog Number 448606
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There were multiple 510k numbers reported to be involved.The information for the additional 510k is as follows: g4.Pma / 510(k)#: k020322, k021954, k022172, k023273, k023301, k024152, k030677, k031306, k031679, k031679, k032131, k033784, k033907, k040006, k040106, k040716, k050555, k051689, k053241, k060214, k060217, k060218, k060218, k060493, k082538, k082852, k082913, and k131331.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigatm 1119779-2025-00034ion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using the bd phoenix¿ pmic/id-106 an unspecified number of patients from a blood culture and qc isolates (staphylococcus aureus) were misidentified as staphylococcus epidermidis.No health impact or consequence reported.
 
Additional Manufacturer Narrative
Investigation summary: this complaint is for misidentification of staphylococcus aureus as staphylococcus epidermidis when using phoenix panel pmic/id-106 (catalog number 448606) batch number 4317030.The customer did not return panels, isolates or phoenix generated lab reports for the investigation.To investigate, retention panels of the complaint batch and control panels from the same material but a different batch number were tested using in house isolates s.Aureus a25923 on a phoenix m50 instrument and evaluated for identification results.All panels tested returned the correct identification.This complaint is not confirmed.The batch history record was satisfactory, and no quality notifications were generated during manufacturing and inspection.Complaint trending was performed, and no trends were identified associated with this defect.As no trends were identified, no corrective actions are slated at this time.Bd will continue to monitor for trends and take action as necessary.Please continue to communicate any additional concerns.
 
Event or Problem Description
It was reported while using the bd phoenix¿ pmic/id-106 an unspecified number of patients from a blood culture and qc isolates (staphylococcus aureus) were misidentified as staphylococcus epidermidis.No health impact or consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHOENIX¿ PMIC/ID-106
Common Device Name
SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION
Manufacturer (Section D)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key21313449
Report Number1119779-2025-00034
Device Sequence Number15491565
Product Code LON
UDI-Device Identifier30382904486069
UDI-Public(01)30382904486069(11)241112(17)251117(10)4317030
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
SEE H11
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/17/2025
Device Catalogue Number448606
Device Lot Number4317030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2025
Supplement Date Received by Manufacturer04/23/2025
Initial Report FDA Received Date02/05/2025
Supplement Report FDA Received Date05/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-