| Model Number |
SL-0008 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hematoma (1884)
|
| Date of Event |
04/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As part of capa 0032, a retrospective review was conducted to analyze the failure to submit this mdr within the required 30-day timeframe.The review identified inadequacies in the complaint handling procedure, including the absence of standardized definitions for reportable events aligned with fda regulations.This lack of harmonization significantly contributed to the misidentification of this complaint as a reportable mdr.Additionally, the complaint handling unit did not have a structured workflow to guide decision-making when initial information was insufficient, further impacting timely and accurate reporting.
|
| |
|
Event or Problem Description
|
|
The physician applied siteguard to the breast pocket, in the "gutter area," which refers to the lateral edge of the pocket near the underarm.The siteguard solution remained in the pocket for approximately one minute before being suctioned out.It was reported that the patient developed a hematoma following the surgery.
|
| |
|