|
Sterile samples from the reported lot, including the actual devices, were not returned for visual review or analysis.Without the actual reported devices, photos of the devices, or detailed information on the method used to remove the device from the packaging, preoperative preparation of the device, or the surgeon¿s technique, a definitive root cause cannot be determined at this time.Review of labeling 03-5401r3: (vlm-2004).Contraindications: quill¿ monoderm¿ device is not to be used where extended approximation of tissue under stress is required and is not to be used in conjunction with or for fixation of prosthetic devices (e.G.Heart valves or synthetic grafts) that are non-absorbable in nature.Warnings: the loop of the quill¿ device for wound closure should be deployed by users familiar with surgical procedure, techniques and tissues.Physicians should consider the quantity and quality of tissue in which the loop will be anchored.The use of this suture may be inappropriate in highly vascularized tissue or fragile tissue that cannot withstand potential further tightening/constriction within the loops.Users should be familiar with surgical procedures and techniques involving absorbable sutures before employing quill¿ monoderm¿ device for wound closure, as risk of wound dehiscence may vary with the site of application and the suture material used.Physicians should consider the in vivo performance (under actions section) when selecting a suture for use in patients.The use of this suture may be inappropriate in elderly, malnourished or debilitated patients, or in patients suffering from conditions which may delay wound healing.The safety and effectiveness of quill¿ monoderm¿ device have not been established for use in fascial closures (including abdominal wall, thoracic and extremity fascial closures), gastrointestinal anastomoses, cardiovascular tissue, neural tissue, osseous tissue, tendinous tissue, ophthalmic surgery or for use in microsurgery.Precautions: the quill¿ device contains a loop end and barbs to anchor tissues and does not require knots to approximate opposing edges of a wound.Tying of knots on the barbed section of the material will damage the barbs and potentially reduce their effectiveness.For the device to function properly, quill¿ device must first be anchored in the tissue by the deployment of the loop around robust tissue, then with the other subsequently engaged in the tissue by the barbs.Additionally, when completing placement of the barbed segment, an additional backstitch or bite of tissue lateral to the end of the incision is required to lock the quill¿ device in place.Avoid contacting the quill¿ monoderm¿ device and associated needles with other materials (e.G.Surgical gauze, drapes, etc.) in the surgical field to prevent ensnaring on the barbs.If the barbs catch, carefully pull the material in the opposite direction of the needle to disengage it from the barbs.Care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments, such as needle holders and forceps.Do not pull the quill¿ monoderm¿ device out of the package by the needles as this can cause the barbs to catch on one another.Do not attempt to remove memory in the polymer by running fingers down the suture material as this can damage the barbs.Infections, erythema, foreign body reactions, transient inflammatory reactions and in rare instances dehiscence are typical or foreseeable risks associated with any suture and hence are also potential complications associated with the quill¿ device.Acceptable surgical practice should be followed with respect to drainage and closure of infected wounds.Reshaping needles may cause them to lose strength and be less resistant to bending and breaking.Users should exercise caution when handling surgical needles to avoid inadvertent needle sticks.Discard used needles in designated biohazard ¿sharps¿ containers.
|