| Catalog Number |
221700056 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, after the ceramic liner was extracted, the scrub nurse noticed the missing tip on the 56 extractor body.Broken tip was retrieved after ceramic liner was extracted.Surgeon was still able to use the extractor to complete the case.There was no patient consequence and no surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary according to the information received, " after the ceramic liner was extracted, the scrub nurse notice missing tip on the 56 extractor body.Broken tip retrieved after ceramic liner was extracted." the product was not returned to medtech orthopaedics, however photos were provided for review.See attachment ¿pc-001793402 re broken pinnacle ab extractor body 56 ref.221700056 at rhch.¿ the photo investigation revealed that ab extractor body 56mm (221700056) had broken tip.The observed condition of the device was consistent with a component failure that may have been caused by exposure to unintended forces.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the ab extractor body 56mm would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to unintended use error caused or contributed to event and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|