|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The scalp clip (id 201037) was returned for evaluation.Device history record (dhr) - the reported product lot number was noted that a report was opened, and disposition was to do 100% inspection on material.This issue was not caught during the inspection therefore we have notified supplier.No other anomalies were identified that occurred during the manufacture or packaging of the device.Failure analysis - upon visual inspection of the disposable's packaging, the failure evaluator confirmed the presence of a foreign object inside the pack.The complaint was confirmed.Root cause analysis - a potential root cause is operator error, workstation not clean.
|